USE DATA FROM PERIOD: 252.845 367.000 CYLINDER POSITIONS (UNKNOWN=0, CO2 AND O2 KNOWN=1, CO2 ONLY KNOWN=2, DO NOT USE=8, NOT RUN=9): 1 0 2 9 3 1 4 1 5 0 6 0 7 1 8 1 9 0 10 0 11 0 12 1 13 0 14 0 15 0 16 9 NOTES: Unknowns Pos ID Description 1 D269489 Next WAO WSS3 5 ND29110 former WAO WT / used in Picarro O2 tests 6 D269491 Next WAO WSS4 9 D753898 OXYFLUX 10 D548325 OXYFLUX 11 D548332 OXYFLUX 13 D073408 Next WAO WSS2 / former HBA WSS or TT 14 D743657 Next WAO WSS1 / former HBA WSS or TT 15 D548331 OXYFLUX Knowns Pos ID SIOCO2 GMDCO2 SIOO2 Description 3 D089507 428.087 428.087 -565.460 CRAM WSS5 4 D801299 381.475 381.475 -486.054 CRAM WSS1 7 D073409 398.707 398.707 -658.410 CRAM WSS2 8 D073419 440.065 440.065 -926.360 CRAM WSS3 12 D073418 413.092 413.092 -782.739 CRAM WSS4 Working Tank D258968 - Running 4x next WAO WSS, 2 of which are also former HBA WSS or TT; 1x former WAO WT / Picarro O2 test cylinder; 4x OXYFLUX. - Running CRAM WSSes as knowns for CO2 and O2 results. - NB: have not resolved SIO vs GMD CO2 scale - CRAM WSSes currently reporting only a single scale (which?!?). - Lines 13 and 14 have been swapped, at the point of the QC jumpers - investigating line 14 zig zag problem. - Conclusion: Unclear; do again with current set up.